Medical imaging breakthrough could transform cancer and arthritis diagnosis
30 September 2024
Publish date: 22 November 2023
Issue Description:
A software error has been identified in the CamAPS FX app version 1.4(174) and earlier versions, where the countdown to bolus delivery was initiated twice. This error is extremely rare but may lead to repeated delivery of a standard insulin bolus when initiated from the app.
In order address this error, the software has been updated and an additional control measure, preventing the delivery of a bolus within 1 minute of the previous bolus, has been added. These software changes have been added to App version 1.4(175).
There was no issue with Dana Insulin Pump or mylife YpsoPump.
What you need to do?
The following guidelines should be followed:
Was this page helpful? Let us know